E-97
C. W. Lam, K. P. Soh, S. F. Lim. Singapore General Hospital. Singapore. Singapore.
Aim: The purpose of this study was to evaluate Beekman Coulter UniCel DxI 800 (DxI) for linearity, analytical sensitivity and imprecision for selected chemistries under routine laboratory conditions. We also compared the DxI against another immunoassay analyzer. Roche Klccsys 2010.
Methods: The chemistries of ferritin. Folate, and vitamin B12 are evaluated. The within-run (n=21) and between-day (n=10) imprecision was determined hy the use of three levels of independent commercial control material (Biorad Lyphochek®). Between-day...